Truncations and in silico Docking to Enhance the Analytical Response of Aptamer-Based Biosensors

10 May 2024, Version 1
This content is a preprint and has not undergone peer review at the time of posting.

Abstract

Aptamers are short oligonucleotides capable of binding specifically to various targets (i.e., small molecules, proteins, and whole cells) which have been introduced in biosensors such as in the electrochemical aptamer-based (E-AB) sensing platform. E-AB sensors are comprised of a redox-reporter-modified aptamer attached to an electrode that undergoes, upon target addition, a binding-induced change in electron transfer rates. To date, E-AB sensors have faced a limitation in the translatability of aptamers into the sensing platform presumably because sequences obtained from Systematic Evolution of Ligands by Exponential Enrichment (SELEX) are typically long (> 80 nucleotides) and that obtaining structural information remains time and resource consuming. In response, we explore the utility of aptamer base truncations and in silico docking to improve their translatability into E-AB sensors. Here, we first apply this to the glucose aptamer, which we characterize in solution using NMR methods to guide design and translate truncated variants in E-AB biosensors. We further investigated applicability of the truncation and computational approaches to five other aptamer systems (vancomycin, cocaine, methotrexate, theophylline, and ochratoxin A) from which we derived functional E-AB sensors. We foresee that our strategy will increase the success rate of translating aptamers into sensing platforms to afford low-cost measurements of molecules directly in undiluted complex matrices.

Keywords

Aptamer docking
Aptamer truncations
Electrochemical aptamer-based biosensors
glucose biosensor
Electrochemical impedance spectroscopy
NMR spectroscopy

Supplementary materials

Title
Description
Actions
Title
Supporting Information
Description
Docking results and predicted structures for all aptamers, experimental methods for E-AB biosensors fabrication and characterization, E-AB biosensors impedimetric results and NMR characterization of glucose aptamer with its variants. The molecular docking results of this study are available from the corresponding author upon reasonable request.
Actions

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .
This site is protected by reCAPTCHA and the Google Privacy Policy [opens in a new tab] and Terms of Service [opens in a new tab] apply.