A high macro molecular ghost in SEC analysis of conjugate polymer

29 November 2023, Version 2
This content is a preprint and has not undergone peer review at the time of posting.

Abstract

SEC (size exclusion chromatography) analyses of conjugated polymers (polythiophene and polyphenylene) were carried out. When a semimicro analytical column was employed for the measurement of the molecular weight, an extremely large molecular weight peak (>100000 Da) was observed. Preparative SEC of each fraction and the further analysis revealed that such a high-molecular-weight peak was only found in the analysis with a semimicro column and found that the ghost peak was derived from low-molecular-weight fraction (M < 4000), which would form aggregation to higher molecular weight of earlier eluate in SEC.

Keywords

SEC
Conjugate polymer
polythiophene
polyphenylene

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .