Q-band EPR cryoprobe

06 July 2023, Version 1
This content is a preprint and has not undergone peer review at the time of posting.

Abstract

Following the success of cryogenic EPR signal preamplification at X-band, we present a Q-band EPR cryoprobe compatible with a standard EPR resonator. The probehead is equipped with a cryogenic ultra low-noise microwave amplifier and its protection circuit that are placed close to the sample in the same cryostat. Our cryoprobe maintains the same functionality and compatibility as the commercial instrument allowing high-power pulsed EPR experiments of typical samples. The performance of our setup is benchmarked using pulsed EPR and ENDOR experiments revealing a significant sensitivity improvement, which reduces the measurement time by a factor of about 40x at low temperature.

Keywords

EPR
cryoprobe
Q-band
noise

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .