Abstract
The development of microfluidic systems for biological assays presents challenges, particularly in adapting traditional optical absorbance assays to smaller volumes or to microfluidic formats. This often requires assay modification or translation to a fluorescence version, which can be impractical. To address this issue, our group has developed the µChopper device, which uses microfluidic droplet formation as a surrogate for an optical beam chopper, allowing for lock-in analysis and improved limits of detection with both absorbance and fluorescence optics without modifying the optical path length. Here, we have adapted the µChopper to low-cost optics using a light-emitting diode (LED) source and photodiode detector, and we have fabricated the pnuematically valved devices entirely by 3D printing instead of traditional photolithography. Using a hybrid device structure, fluidic channels were made in polydimethylsiloxane (PDMS) by moulding onto a 3D-printed master then bonding to a prefabricated thin layer, and the pneumatic layer was directly made of 3D-printed resin. This hybrid structure allowed an optical slit to be fabricated directly under fluidic channels, with the LED interfaced closely above the channel. Vacuum-operated, normally closed valves provided precise temporal control of droplet formation from 0.6 to 2.0 Hz. The system was validated against the standard plate reader format using a colorimetric fructosamine assay and by quantifying fructosamine in human serum from normal and diabetic patients, where strong correlation was shown. Showing a standard benefit of microfluidics in analysis, the device required 6.4-fold less serum volume for each assay. This µChopper device and lower cost optical system should be applicable to various absorbance based assays in low volumes, and the reliance on inexpensive 3D printers makes it more accessible to users without cleanroom facilities.
Supplementary materials
Title
Supporting info: Droplet-based µChopper device with a 3D-printed pneumatic valving layer and a simple photometer for absorbance based fructosamine quantification in human serum
Description
As mentioned in the main manuscript, supplemental material for Figure S-1 and Figure S-2 is provided, showing the photodiode slit characterization as well as path length measurement and the LED intensity characterization respectively.
Actions