Trace Measurements of Ethylene Oxide using Cavity-enhanced Absorption Spectrometry near 3066 cm-1

28 January 2022, Version 1
This content is a preprint and has not undergone peer review at the time of posting.

Abstract

Ethylene oxide (EtO) is a key carcinogen that is widely used in chemical manufacturing and biotechnology industries. Recent work has suggested that permissible exposure limits for EtO be reduced from 1 – 5 ppm to sub-ppb levels. Such new standards will require new methodologies that are capable of measuring EtO with the requisite precision. In this paper, we demonstrate a new analyzer based on cavity-enhanced absorption spectrometry that utilizes a broad EtO absorption feature near 3066 cm-1. A fit function is developed that includes water, methane, and EtO absorbances and accounts for absorption both inside and outside the cavity. A methane standard is used to determine the cavity gain factor, and the EtO absorbance spectrum is empirically determined. The final system shows excellent linearity from 0 – 909 ppb EtO (R2 ~ 0.9999) with a measurement precision of better than ±1 ppb (1-sigma, 60 seconds) that improved to ±0.5 ppb (1-sigma, 15 minutes). Deliberate ambient EtO releases demonstrate the instrument’s utility in rapidly detecting hazardous conditions. Further work will include improving the measurement precision and directly comparing the system to EPA Method TO-15.

Keywords

ethylene oxide
cavity ringdown
cavity-enhanced absorption spectroscopy

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .
This site is protected by reCAPTCHA and the Google Privacy Policy [opens in a new tab] and Terms of Service [opens in a new tab] apply.