Using voltammetry augmented with physics-based modeling and Bayesian hypothesis testing to identify analytes in electrolyte solutions

23 August 2021, Version 3
This content is a preprint and has not undergone peer review at the time of posting.

Abstract

Voltammetry is a foundational electrochemical technique that can qualitatively and quantitatively probe electroactive species in solutions and as such has been used in numerous fields of study. Recently, automation has been introduced to extend the capabilities of voltammetric analysis through approaches such as Bayesian parameter estimation and compound identification. However, opportunities exist to enable more versatile methods across a wider range of solution compositions and experimental conditions. Here, we present a protocol that uses experimental voltammetry, physics-driven models, binary hypothesis testing, and Bayesian inference to enable robust labeling of analytes in multicomponent solutions across multiple techniques. We first describe the development of this protocol, and we subsequently validate the methodology in a case study involving five N-functionalized phenothiazine derivatives. In this analysis, the protocol correctly labeled solutions each containing 10H-phenothiazine and 10-methylphenothiazine from both cyclic voltammograms and cyclic square wave voltammograms, demonstrating the ability to identify redox-active constituents of a multicomponent solution. Finally, we identify areas of further improvement—such as achieving greater detection accuracy—and future applications to potentially enhance in situ or operando diagnostic workflows.

Keywords

Cyclic voltammetry
Cyclic square wave voltammetry
Analyte identification
Bayesian inference
Physics-based modeling
Phenothiazine

Supplementary materials

Title
Description
Actions
Title
Fenton-Jr-&-Brushett SI
Description
Supplementary Information for the main text. Discusses methods and analyses in greater detail.
Actions

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .
This site is protected by reCAPTCHA and the Google Privacy Policy [opens in a new tab] and Terms of Service [opens in a new tab] apply.