Performance of a FRET-Based Point-of-Care Immunoassay for the Quantitation of Fecal Calprotectin

18 February 2021, Version 1
This content is a preprint and has not undergone peer review at the time of posting.

Abstract

A fast (~5 min), time-resolved fluorescence resonance energy transfer based immunoassay (Procise FCP™) was developed for the point-of-care quantitative detection of fecal calprotectin (FCP) using 15 mg of fecal specimen eluted in collection fluid from the Procise Stool Collection Device™. Studies were performed to characterize analytical performance of the Procise FCP assay on the ProciseDx™ analyzer.


The Procise FCP assay showed good analytical performance with respect to sensitivity, specificity, linearity, and precision suitable for routine clinical use in a point-of-care setting as well as excellent analytical agreement with a current commercial FCP measurement method.


Results indicate that the Procise FCP assay is sensitive, specific, and precise yielding results in less than 5 minutes. This indicates the Procise FCP assay is useful for obtaining fast and accurate FCP quantitation, thus avoiding delays inherent to current methods and enabling immediate clinical assessment to be made during a single patient visit.

Keywords

Crohn's disease
Fecal calprotectin
FRET
Inflammatory bowel diseases
Lumiphore
Point of care test
Ulcerative Colitis

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .